Benchtop temperature humidity test chamber support customer reliability testing with small size and amazing testing performance. When the temperature and humidity rising, when the benchtop environmental chamber working with a mistake, In order to prevent the test sample from being damaged by condensed water, The Wewon Benchtop temperature humidity test chamber added additional feature to prevent frosting during test period. The application of benchtop temperature humidity chamber : Mini samples test, such as micro semiconductor chip, IC lead frame; Stability test of new products in food, ecology, textile fiber and medicine industry.
The Wewon bench-top type temperature humidity chamber with a convenient touch screen control system, The test record choose SD card memory terms, There also include RS 485, RS 232C, and Ethernet communication port which will is convenient to connect with your computer system or remote control system. The free PC link software will be supplied when customer ordred the benchtop environmental test chamber from Wewon Environmental Chambers Co., Ltd.

Equipment Model | WEW-22L-TA | WEW-22L-TD | WEW-64L-TA | WEW-64L-TD |
Internal Dimensions (W*H*D) | 300*300*250 | 300*300*250 | 400*400*400 | 400*400*400 |
External Dimensions (W*H*D) | 450*690*795 | 450*690*1000 | 550*790*1015 | 550*790*1170 |
Temperature Range | -40℃~150℃ | -70℃~150℃ | -40℃~150℃ | -70℃~150℃ |
Cooling Rate | 20℃~-40℃ within about 60 minutes | 20℃~-70℃ within about 90 minutes | 20℃~-40℃ within about 70 minutes | 20℃~-70℃ within about 100 minutes |
Heating Rate | 40℃~150℃ Within 60 minutes | -70℃~150℃ Within about 70 minutes | 40~150℃ Within about 70 minutes | -70℃~150℃ Within about 80 minutes |
Temperature Fluctuation | ±0.3℃ | ±0.3℃ | ±0.3℃ | ±0.3℃ |
Temperature Distribution | ±1.0℃(-70℃~100℃) | |||
±2.0℃(100.1℃~150℃) | ||||
Humidity range | 30~98%RH | 30~98%RH | 30~98%RH | 30~98%RH |
Humidity Variation | ±1.0%RH | ±1.0%RH | ±1.0%RH | ±1.0%RH |
Humidity Distribution | ±2.0%RH | ±2.0%RH | ±2.0%RH | ±2.0%RH |
Internal Material | Stainless steel plate (SUS304) | |||
External Material | Cold rolled steel sheet / powder coating | |||
Thermal Insulation Material | Polyurethane Board + Glass Wool | |||
Fan System | lmpeller Fan︱Wewon Benchtop Temperature Humidity Test Chamber | |||
Compressor | Rolling (frequency conversion type) | Rolling (frequency conversion type) binary freezing | Rolling (frequency conversion type) | Rolling (frequency conversion type) binary freezing |
Condenser | Air-cooled | Air cooled | Air-cooled | Air cooled |
Refrigerant | R23,R507 | R23,R507 | R23,R507 | R23,R507 |
Evaporator | Air-cooled Plate Fin Cooler | |||
The Heater | Nickel-Chromium Heating Wire | |||
Humidifier | Stainless steel cartridge heater | Stainless steel cartridge heater | Stainless steel cartridge heater | Stainless steel cartridge heater |
Net Weight | 80kg | 100kg | 140kg | 150kg |
Ambient Temperature | +5~35℃︱Testing Hole: Φ50mm, 2pcs︱Shelves Plate, 2 pcs | |||
Safety Specification | Cut off the control circuit to prevent excessive temperature rise, overload relay, chiller protection circuit, leakage cut-off machine, external alarm terminal, water shortage alarm. |
Benchtop temperature and climatic test chambers simulates your samples testing work with temperature range -70C +150C and humidity range 30%~95% for product performance evaluation purpose ! The benchtop test chamber include one powerful France Tecumseh compressor, and two-kilowatt heating device that has optimum cooling and heating performance. The testing room 300*300*250mm (W*D*H) of Wewon benchtop environmental chamber keeps the flexibility to accommodate both mini type and large size test samples.
According with the requirements of today’s science laboratory environments, The structure of Wewon bench-top type temperature & humidity chamber incorporates a compact exterior that maximizes the inside testing room. If continuous testing is carried out, repeated heating and cooling test will reduce the performance of benchtop temperature humidity chamber. Therefore, We recommend that diversified sample tests should be given priority in the testing period. During the operation of the climate test chamber, Please inject distilled water into the water tank. Because tap water will shorten the service life of the heater.

● | The pull-out water supply bucket located at the lower end of the main body is detachable, free to move and convenient to replenish water | Convenient Water Supply | Product Characteristics | |
● | A highly reliable dry and wet ball control method is used to obtain precise test results and to design long-term tests. | Dry And Wet Ball Sensors | ||
● | AC110V/220V power supply low power consumption design, general household socket can be easily installed. | Low Power Consumption | ||
● | BLDC motor is adopted to realize the low noise below 55dB, which is suitable for indoor use, just like the quiet requirement of laboratory and office. | Low Noise | ||
● | The Ethernet port remote technology is used to remotely control the test equipment, monitor and send emails to PC and mobile phone in real time, thus enhancing the convenience of users︱Benchtop Environmental Chamber Technical Parameters | Easy to Connect | Remote Network System | |
Send Email | ||||
Manager Function | ||||
● | In case of system error, the contents of each error will be displayed on the monitor with photos for the fastest confirmation and response | Convenient for Checking System errors | ||
● | In order to prevent samples from being damaged by condensation when temperature and humidity rise and errors occur, the function of preventing | Prevent Sample Condensation Control | ||
● | In order to prevent dew condensation, the speed of change at the time of temperature and humidity rise is suppressed, and the set temperature reaches → waits for the set temperature for 5 minutes → the program starts with humidification. | Humidification Delay Operation | ||
● | After the end of the test, the humidified water in the laboratory was drained → the test chamber was lowered to 23 ° C → the test was completed after waiting. | The Test Ended in Operation | ||
● | Protect the safety of the machine and the user by cutting off the overheating that occurs during use (applicable to electronic and electrical mechanical triple-cut devices) and overcurrent. | Cut Off Overheating And Overcurrent | ||
● | In smaller sizes, it can also be used in a small space like a laboratory or an office. | Compact | ||
● | Outputs such as communication and operation, alarms, and time signals (RUN, Alarm, Time Singnal) desired by the user are provided by relay contacts. | Service Port | ||
● | A cable port that can be connected to the power supply of the sample and a cable for measurement is mounted on the side of the main body. | Cable Port |
The optimum quantity of refrigerant is controlled by adjusting the electronic expansion valve. Adjust the wasted calorific value and refrigerating capacity to make sure that the Wewon benchtop temperature humidity test chamber will save energy. For humidity testing needs of customers, the benchtop temp humidity test chamber can achieve automatic water supply function. When the pressure of the purified water inlet is more than 3kg/cm2, the pressure regulating valve shall be installed. No relief valve is required when the pressure at the purified water inlet is below 3kg/cm2. If automatic water supply is required, it needs to be specified when ordering to us.
Look After Your Testing With This Superb Benchtop Environmental Chamber Recommendation